Estimation of the Ablated Area Size based on Local Conduction Velocity: Simulations and Animal Experiments

Jimena Siles Paredes1, João Salinet2, Stefan Pollnow3, Olaf Doessel4, Ilija Uzelac5
1Biomedical Engineering, Engineering, Modelling and Applied Social Sciences Centre, Federal University of ABC, 2Biomedical Engineering, Engineering, Modelling and Applied Social Sciences Centre, 3Karlsruhe Institute of Technology (KIT) Institute of Biomedical Engineering, 4Institute of Biomedical Engineering, Karlsruhe Institute of Technology, 5Georgia Institute of Technology


Abstract

Aims: This study aimed to validate a method termed the circle method to estimate local conduction velocity (CVL) magnitude and direction based on optical mapping action potentials (OAPs) to characterize the ablated area. Method: Simulations of the wavefront propagation (planar and curve) were done using the Fenton-Karma 3V model, where an ablated area was modeled by changing the conductivity parameter. Additionally, radio-frequency ablation (RFA) was performed in isolated atria of rats and isolated guinea pigs’ hearts for different RFA durations ranging from 0.5 s to 14.5 s. Optical mapping imaging was done using voltage-sensitive dyes to acquire OAPs and construct local activation time (LAT) maps. For each pixel in a selected region with the ablated area inside, CVL’s magnitude and direction were calculated along a circular disposition spanning from 0◦ to 180◦ outlining a circle of specific radius. In this fashion, the maximal CVL corresponds to the direction parallel to the propagating wavefront, allowing estimation of the actual CVL located at the angular distance of 90◦. Results: CVL at the center of the ablation area showed a logarithmic curve increasing with larger radii, suitable for estimating the ablated area size. From the CVL maps, distinct regions of substantially lower CVLs (from 2.95 to 2.40 cm/s for simulation, 70 to 30 cm/s for the atrium, and 120 to 100 cm/s for the ventricle) showed the ablated area’s location. Maps of CLV directions around the ablated area show characteristic deviation symmetric regarding the propagating wavefront allowing for more accurate delineation of ablated area. Conclusions: The method is validated for different ablation geometries and sizes, wavefront curvatures, and different animal experiments. Estimation of ablated area’s size and its location could be performed without prior knowledge of the wavefront propagation direction.